Point of Product
Specification
Materials of Construction
All materials are FDA-listed materials per 21-CFR
Membranes
|
Double layer PTFE
|
Support and Drainage Layers
|
Polypropylene, oxidation resistant
|
Core, Cage and Endcaps
|
Polypropylene, oxidation resistant
|
Adapters
|
Polymeric, encapsulated in oxygen-resistant polypropylene
|
O-rings
|
Silicone elastomer
1
|
1
Other materials available on request
Removal Ratings
Air/Gases
|
<0.003 µm, particulate
2
|
Liquids
|
0.2 µm, sterilizing
3
|
2 By NaCl CNC particle analysis. Also retains 100 % of Brevundimonas diminuta bacteria, bacteriophage and virus aerosols.
3
Lot samples of assemblies retain >10
7
cfu/cm
2
of
Brevundimonas diminuta in liquid per mod. ASTM F838-83 and FDA guidelines.
Configurations (AB Code 7) 4
Single open ended
|
Double 226 O-ring adapter
|
Finned end, bayonet lock
|
Dimensions (Nominal)
Lengths
|
127 mm (5 in.), 254 mm (10 in.), 508 mm (20 in.), 762 mm (30 in.), 1016 mm (40 in.)
|
Diameter
|
70 mm (2.75 in.)
|
0.8 m2 (8.6 ft2) per 254 mm (10 in.) element |
0.4 m2 (4.1 ft2) per 127 mm (5 in.) element |
Temperature
|
Maximum Differential Pressure
|
20°C
|
5.3 bar (77 psid)
|
80°C
|
4.1 bar (60 psid)
|
Typical Continuous Air Service Life 6
12 months to 60°C
|
Typical Vent Service Life
6 months to 80°C
|
Typical Vent Service Life
Autoclave/Streaming
|
Pressure and temperature
|
Cumulative steaming time
8,9
|
Maximum Forward Steaming Conditions
8,9
|
1.0 bard (14.5 psid) at 125 °C
0.3 bard (4.4 psid) at 142 °C |
20 hrs (1 hr cycles)
165 hrs (1 hr cycles) |
Maximum Forward Steaming Conditions
8,9
|
0.5 bard (7.3 psid) at 125 °C
|
20hrs (1 hr cycles)
|
8 Lab tests to establish multi-cycle resistance. Filters should be qualified in actual use. Contact Pall for recommended procedures.
9 Excursion conditions only
Aqueous Extractables (20°C)
<5 mg per 254 mm (10 in.) element after 1 hour steam at 125 °C
|
Biological Tests
|
Meets USP Biological Reactivity, In Vivo, for Class VI-121 °C plastics
|
Effluent Quality Tests
9
|
Meets Cleanliness per USP Particulates in Injectables. Non-Fiber-Releasing per 21CFR Non-Pyrogenic per USP Bacterial Endotoxins (<0.25 EU/mL) Meets Oxidizables and pH per USP packaged waters |
Steam Resistance
|
Lot samples multicycle autoclave challenged
|
Integrity Test Values per 254 mm (10in) Catridge 10
Forward Flow (air test gas)
|
60:40 IPA:Water wet (v.v) ≤ 15 mL/min at 1040 mbar (15 psi)
|
Water Intrusion
|
≤ 0.33 mL/min at 2500 mbar (36 psi)
|
Ordering Information
Part
Number 11
|
Nominal
Length
|
AB Style Code
10
|
AB05PFR2PVH4
|
127
mm (5 in.)
|
2 |
AB1PFR7PVH4
|
254
mm (10 in.)
|
7 |
AB2PFR7PVH4
|
508
mm (20 in.)
|
7 |
AB3PFR7PVH4
|
762
mm (30 in.)
|
7 |
AB4PFR7PVH4
|
1016
mm (40 in.)
|
7 |
11 P at end of part number indicates optimized and qualified for Pharmaceutical use including encapsulated stainless steel reinforcing adapter ring for steaming in situ, 100% integrity testing and Certificate of Test provided. V indicates optimized for vent/air/gas service. H4 suffix=Silicone O-rings (standard). See Appendix for available alternate O-ring polymers and codes.
12 AB05 standard with Code 2 flat cap, other lengths Code 7 with finned cap. Both with double 226 O-rings.
Note: For sizing filters or other information, please contact your local Pall distributor.
Specifications and Availability: The information provided is a guide to the part number structure and possible options. Product availability may be subject to change without notice. All specifications are nominal. This literature was reviewed for accuracy at the time of publication. For current information on the products and test methodologies, consult your local Pall distributor.
Type
Filter Catridges
Use
Air, Gas, and Vent Filtration