The Octet® R4 System Makes Biologics Development Affordable and Flexible
Point of Product
Benefits
1. Fast Results :
The Octet® R4 system has four parallel, independent channels for increased flexibility and faster results.
2. High Sensitivity, Moderate Throughput, Low Cost :
The Octet® R4 system has four parallel, independent channels for increased flexibility and faster results.
3. Reduced Assay Time and Maintenance Cost :
The Octet® platform provides faster time to results and process economy when compared to other technologies due to its fluidics-free Dip and Read format. Additionally, crude sample compatibility saves valuable time by eliminating the need for sample pre-treatment, and non-destructive testing conserves precious samples for other assays.
4. Field Upgradeable within a Single Visit :
Choose when to upgrade to an 8-channel Octet® system for higher throughput and full GMP compliance. Upgrades are performed on-site within a single visit by trained Sartorius engineers without the hassle of down time or system trade-in.
5. Sample Plate Cooling for Temperature-Sensitive Proteins
The sample plate temperature can be controlled from 15–40°C, which enables reliable kinetic determinations at different temperatures for temperature-sensitive proteins. Additional advantages of sample cooling include the ability to rapidly determine binding rate constants at multiple temperatures to extrapolate thermodynamic measurements. The Octet® R4 system’s four channels can be used independently to measure samples for screening purposes or in tandem, pairing the sample read with a dedicated reference for high-quality kinetic characterization.
Applications
Kinetic and Affinity Characterization
The Octet® R4 system monitors binding events in real time to calculate on rates (ka), off rates (kd), and affinity constants (KD) by bringing the detection surface directly to the sample, eliminating the need for microfluidics. The sample plate temperature can be controlled from 15–40°C, which enables reliable kinetic determinations at different temperatures for temperature-sensitive proteins. Additional advantages of sample cooling include the ability to rapidly determine binding rate constants at multiple temperatures to extrapolate thermodynamic measurements.
Protein Quantitation in Bioprocessing
Accurately determining the concentration of antibodies and other proteins is critical to the selection of cell lines for development and the optimization of antibody production. The Octet® R4 system directly measures the presence of specific proteins and other molecules in solution with minimal interference from complex matrices with reproducible concentrations determined in as little as 2 minutes per sample using a simple, one-step assay. Rapid turnaround time of results improves process efficiencies and productivity that cannot be achieved with othermethods such as HPLC and ELISA.
Specification
Detection
Technology
|
Bio-Layer
Interferometry (BLI)
|
Biosensor
Type
|
Disposable,
single-use fiber optic biosensors with optional reuse by regeneration and/or
re-racking in the biosensor tray
|
Data
Presentation
|
Graphs
displaying real-time kinetic binding sensorgrams, fitted result plots, and
residuals of fits Concentration data analysis including calibration curves and output of tabulated concentration data Tabulated kinetic data Epitope binning and cross-blocking matrices and trace overlays |
Sample
Type
|
Proteins,
antibodies, peptides, DNA, RNA, liposomes, viruses and VLPs in various media
including serum, buffers containing DMSO, periplasmic fractions, untreated cell
culture supernatants, and crude cell lysates
|
Number
of Spectrometers
|
4 |
Maximum
Simultaneous Reads
|
4 |
Data
Collection Rate
|
2,
5 or 10 Hz
|
Sample
Position and Format
|
1 standard 96-well, black, flat-bottom microplate |
Sample
Volume
|
180–220
µL/well Nondestructive testing |
Orbital
Flow Capacity
|
Static
or 100–1500 rpm
|
Analysis
Temperature Range
|
15°C
to 40°C, in 1°C increments
|
Throughput
|
Up
to 4 assays in parallel
|
Molecular
Weight Detection
|
>150 Da |
Analysis
Time per Sample
|
Real-time
kinetic binding experiments from 5 minutes to 4 hours
|
Association
Rate Constant (ka)
|
101 to 107 M-1 s-1 |
Dissociation
Rate Constant (kd)
|
10-6 – 0.1 s-1 |
Affinity
Constant (KD)
|
1
mM – 10 pM
|
Baseline
Noise (RMS)
|
<3.0
pm
|
Baseline
Drift
|
≤
0.12 nm/hour
|
Quantitation
Precision
|
CV
<10%
|
Weight
|
72 kb (32.7 Kg) |
Dimensions (H X W X D)
|
19.5 in X 22 in X 18.2 in
(49 cm X 56 cm X 46 cm) |